Panasonic MDF-U5312 Series Troubleshooting Manual
TROUBLE SHOOTING
If the unit malfunctions, check out the following before calling for service.
Malfunction | Check/Remedy |
The chamber is not cooled at all | - The circuit breaker of power source is active
- The voltage is too low. (In this case, call an electrician.)
- The power switch is not ON.
- The large amount of articles (load) is stored in the chamber at one time.
- The freezer is in defrost condition.
|
The cooling is poor | - The ambient temperature is too high.
- The outer door is not closed firmly.
- The large amount of frost is built on the chamber wall.
- The air intake vent is blocked.
- The set temperature is not inputted properly.
- The freezer is in the direct sunlight.
- There is any heating source near the freezer.
- A rubber cap and insulation for the access port are not set correctly.
- You put too many unfrozen articles into the freezer compartment.
|
Noise | - The freezer is not installed on the sturdy floor.
- The freezer is not leveled with the leveling feet.
- There is anything touching the frame.
- The freezer is in the status immediately after start up.
The unit sometimes causes a noise when the chamber temperature is high due to the large load. The noise gets less and less accompanying with the cooling of the chamber. |
Note:
If the malfunction is not eliminated after checking the above items, or the malfunction is not shown in the above table, contact our sales representative or agent.