Page 2
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
1 General information and safety guidelines Validity of the document This document is valid for the subsequent installation of a multi-function footswitch in an ARCADIS Varic system or an ARCADIS Orbic system that has been prepared for the multi-function footswitch option. Requirements...
If work steps must be performed using electrical power, the general safety information according to TD00-000.860.01 must be observed. Page 5 of 20 Siemens AG SPR2-310.812.03.01.02 ARCADIS Varic/Orbic Medical Solutions 09.06 CS PS SP...
(e.g., monitor cable or C-arm cable)) are replaced or if protective conductor connections have been repaired, the protective conductor resistance must be measured. The values must be documented and evaluated in the protective conductor resistance protocol. Page 6 of 20 ARCADIS Varic/Orbic SPR2-310.812.03.01.02 Siemens AG 09.06 CS PS SP...
Page 7
DIN VDE 0751-1/2001-10, Fig. C2. Pos. 1 = System Pos. 2 = Application part type B (if available) Pos. 3 = Measurement setup (integrated into measuring device) Page 7 of 20 Siemens AG SPR2-310.812.03.01.02 ARCADIS Varic/Orbic Medical Solutions 09.06 CS PS SP...
Mechanical installation and wiring ARCADIS Varic 2Mechanical installation and wiring Opening the covers • Open the rear cover panel of the basic unit. • Disconnect the ground wire from the cover panel. Installing the new footswitch holder • Once the rear cover panel has been removed from the basic unit, you can see three holes in it that are fitted with dummy covers.
Removing the wire from the RNH.3 and RNH.4 relays to D1.X14 • At the RNH (emergency stop relay), remove the RNH.3 and RNH.4 wires. • On the D1 board, detach the D1.X14 plug. Page 9 of 20 Siemens AG SPR2-310.812.03.01.02 ARCADIS Varic/Orbic Medical Solutions 09.06 CS PS SP...
Page 10
(Fig. 6 / p. 10). Attach the ends of the Y cable along the existing cable assemblies using the cable ties included in the acces- sory kit. Page 10 of 20 ARCADIS Varic/Orbic SPR2-310.812.03.01.02 Siemens AG 09.06 CS PS SP...
• Reattach the rear cover panel to the basic unit. • Skip the next section, "ARCADIS Orbic." Continue with the section "Updating the footswitch symbols". Page 11 of 20 Siemens AG SPR2-310.812.03.01.02 ARCADIS Varic/Orbic Medical Solutions 09.06 CS PS SP...
Remove the wire. To do so, cut open the appropriate cable ties. The wire will be replaced with the Y cable included in the accessory kit. Page 13 of 20 Siemens AG SPR2-310.812.03.01.02 ARCADIS Varic/Orbic Medical Solutions 09.06 CS PS SP...
NOTE The RNH.3 and RNH.4 wires are not needed when the 3D recon- struction option is installed. Page 14 of 20 ARCADIS Varic/Orbic SPR2-310.812.03.01.02 Siemens AG 09.06 CS PS SP Medical Solutions...
Reattach the ground wire to the rear cover panel of the basic unit. Make sure the wire has good contact with the cover panel. • Reattach the rear cover panel to the basic unit. Page 15 of 20 Siemens AG SPR2-310.812.03.01.02 ARCADIS Varic/Orbic Medical Solutions 09.06 CS PS SP...
(Fig. 13 / p. 17) here:(4/Fig. 15 / p. 18). NOTE For the ARCADIS Varic and ARCADIS Orbic, the functions of the two footswitch pedals (up to software version VB13C) cannot be changed. From software version VB13C onward, the foot pedal functions will be alterable via the software.
Page 17
Mechanical installation and wiring Fig. 12: "Select Mode" symbol Fig. 13: "Fluoro Pedal" symbol Page 17 of 20 Siemens AG SPR2-310.812.03.01.02 ARCADIS Varic/Orbic Medical Solutions 09.06 CS PS SP...
Page 18
Mechanical installation and wiring Fig. 14: "Release Radiation Pedal" symbol Fig. 15: Multi-function footswitch Page 18 of 20 ARCADIS Varic/Orbic SPR2-310.812.03.01.02 Siemens AG 09.06 CS PS SP Medical Solutions...
After the radiation release stops, the previously selected mode is selected again automatically, and the LED on the corresponding button (PFC, SUB, ROAD or DR) lights up. Page 19 of 20 Siemens AG SPR2-310.812.03.01.02 ARCADIS Varic/Orbic Medical Solutions 09.06 CS PS SP...
• Explain the functions of the multi-function footswitch to the customer. Also, demon- strate how to place the footswitch in the holder. Page 20 of 20 ARCADIS Varic/Orbic SPR2-310.812.03.01.02 Siemens AG 09.06 CS PS SP...